Kit for urea determination
Product name: Kit for urea determination
Method: urease velocity method
Basic index:
amount of specimen: 3 ul |
dosage of R1: 240 ul |
dosage of R2: 60 ul |
dominant wavelength: 340 nm |
sub-wave length: 405 nm |
assay method: velocity method |
direction of reaction: down |
reaction temperature: 37℃ |
scaling mode: Linear |
incubating time of R1+ sample: 300 sec |
reaction time: 90 sec |
Clinical significance: It is for testing of urea content in human serum to test renal function.
Order information:
Product name |
Art.No. |
Packaging specification |
Main technical performance index |
Kit for urea determination(UREA)
|
URE-2070 |
R1: 2 × 40 ml R2: 2 × 10 ml |
◆ blank absorbency≥1.20(340nm) ◆ rate of vacancy change≤0.005A/min ◆ sensitivity for analysis: 8.30mmol/L △A/min≥0.015 ◆ degree of accuracy: relative deviation≤10% ◆ degree of precision: with-in degree of precision: CV≤5% between degree of precision: relative deviation≤8% ◆linearity range0~22.8mmol/L, r≥0.99 |
URE-4070 |
R1: 4 × 40 ml R2: 4 × 10 ml |
||
URE-2060 |
R1: 2 × 80 ml R2: 2 × 20 ml |
||
URE-4060 |
R1: 4 × 80 ml R2: 4 × 20 ml |
||
URE-2080 |
R1: 2 × 80 ml R2: 2 × 20 ml |
||
URE-4040 |
R1: 4 × 40 ml R2: 4 × 10 ml |
◆ Packaging specification with Art. No. of 40 refers to Toshiba 40 type biochemical analyzer used packaging;
◆ Packaging specification with Art. No. of 60 refers to Hitachi 7150, 7060, Shimadzu CL7200, 7300, 8000, Olympus AU800, Toshiba 120FR, Abbott 2000, 8000 biochemical analyzer used packaging;
◆ APackaging specification with Art. No. of 70 refers to Hitachi 170A, 7170, 7080, 7180, 7600P, Olympus AU400, 600, 640, 1000, 2700, 5400, Roche Group P800, Siemens 1650, 1800, 2400 biochemical analyzer used packaging;
◆ Packaging specification with Art. No. of 80 refers to Beckman CX series, Beckman LX20 series, Beckman DXC600/800 series biochemical analyzer used packaging.