Kit for determination of pepsinogenⅡ
Product name: Kit for determination of pepsinogenⅡ
Method: latex-enhanced immunoturbidimetry
Basic index:
amount of specimen: 6 ul |
dosage of R1: 162 ul |
dosage of R2: 30 ul |
dominant wavelength: 700 nm |
sub-wave length: 无 |
assay method: two point end assay |
direction of reaction: forward direction, up |
reaction temperature: 37℃ |
scaling mode: multipoint nolinear |
incubating time of R1+ sample: 300 sec |
reaction time: 300 sec |
Clinical significance: for determination of content of pepsinogenⅡ
Order information:
Product name |
Art.No. |
Packaging specification |
Main technical performance index |
Kit for determination of pepsinogenⅡ(PGⅡ) |
PG2-1170 |
R1: 1 × 27 ml R2: 1 × 5 ml |
◆ blank absorbency≤1.50(700nm) ◆ sensitivity for analysis: 10.0ng/ml △A≥0.020 ◆ degree of accuracy: relative deviation≤10% ◆ degree of precision: with-in degree of precision: CV≤8% between degree of precision: relative deviation≤10% ◆ linearity range: 2.0~100.0ng/ml, r≥0.99 |
PG2-1070 |
R1: 1 × 54 ml R2: 1 × 10 ml |
||
PG2-1060 |
R1: 1 × 54 ml R2: 1 × 10 ml |
||
PG2-2080 |
R1: 2 × 54 ml R2: 2 × 10 ml |
||
PG2-1040 |
R1: 1 × 54 ml R2: 1 × 10 ml |
◆ Packaging specification with Art. No. of 40 refers to Toshiba 40 type biochemical analyzer used packaging;
◆ Packaging specification with Art. No. of 60 refers to Hitachi 7150, 7060, Shimadzu CL7200, 7300, 8000, Olympus AU800, Toshiba 120FR, Abbott 2000, 8000 biochemical analyzer used packaging;
◆ Art.No.尾数为“70”的Packaging specification, 为日立7170A、7170、7080、7180、7600P、奥林巴斯AU400、600、640、1000、2700、5400、罗氏P800、西门子16Z50、1800、2400等生化仪Packaging;
◆ Packaging specification with Art. No. of 80 refers to Beckman CX series, Beckman LX20 series, Beckman DXC600/800 series biochemical analyzer used packaging.
Copyright(C)2015,Shijiazhuang XUBO Chemical Co., Ltd. All Rights Reserved.